Title of article :
Sectional power-law correction for the accurate determination of lutetium by isotope dilution multiple collector-inductively coupled plasma-mass spectrometry
Author/Authors :
Yuan، نويسنده , , Hong-Lin and Gao، نويسنده , , Shan and Zong، نويسنده , , Chun-Lei and Dai، نويسنده , , Meng-Ning، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Pages :
7
From page :
1228
To page :
1234
Abstract :
In this study, we employ a sectional power-law (SPL) correction that provides accurate and precise measurements of 176Lu/175Lu ratios in geological samples using multiple collector-inductively coupled plasma-mass spectrometry (MC-ICP-MS). Three independent power laws were adopted based on the 176Lu/176Yb ratios of samples measured after chemical chromatography. Using isotope dilution (ID) techniques and the SPL correction method, the measured lutetium contents of United States Geological Survey rock standards (BHVO-1, BHVO-2, BCR-2, AGV-1, and G-2) agree well with the recommended values. Results obtained by conventional ICP-MS and INAA are generally higher than those obtained by ID-TIMS and ID-MC-ICP-MS; this discrepancy probably reflects oxide interference and inaccurate corrections.
Keywords :
Lutetium , Sectional power law , multiple collector ICP-MS , Isotope dilution
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy
Serial Year :
2009
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy
Record number :
1680988
Link To Document :
بازگشت