Title of article :
Evaluation of the oscillator frequency of a free running RF generator as a diagnostic tool for inductively coupled plasma-optical emission spectrometry
Author/Authors :
Wiltsche، نويسنده , , Helmar and Moradi، نويسنده , , Farzaneh and Knapp، نويسنده , , Günter، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2012
Pages :
6
From page :
48
To page :
53
Abstract :
The frequency response of a free running RF generator was evaluated for its ability to provide plasma diagnostic information. A small probe coil in the plasma torch housing and a frequency counter were used for frequency measurement. A linear dependence between the RF generator frequency and the flow of a foreign gas (CO2, H2, He, N2—introduced into the central channel of the ICP) was encountered. Moreover, the inductively coupled plasma (ICP) was found to extinguish at a constant, characteristic frequency when the flow of foreign gas was too high, allowing a simple prediction of the maximal tolerable foreign gas flow. Spectroscopic plasma diagnostic techniques (Mg II/Mg I ratio, excitation temperature and rotational temperature) were also investigated but provided little indication of instable plasma conditions due to the addition of foreign gases.
Keywords :
RF generator frequency , Plasma diagnostic , Plasma stability
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy
Serial Year :
2012
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy