Title of article :
Direct solid sampling electrothermal atomic absorption spectrometric determination of toxic and potentially toxic elements in certified reference materials of brown coal fly ash
Author/Authors :
T?r?k، نويسنده , , Peter and ?emberyov?، نويسنده , , M?ria، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2012
Pages :
6
From page :
80
To page :
85
Abstract :
This work describes a method for the direct determination of Bi, Cd, Cr, Ni, Pb, Sb, Sn and Zn in reference materials of brown coal fly ash, using solid sampling electrothermal atomic absorption spectrometry. The use of chemical modifiers was found to be necessary for obtaining reproducible and sufficiently sensitive signals for the analytes under study. The mixture of Pd and Mg(NO3)2 was used for the determination of Bi, Cd, Pb, Sb, Sn and Zn, W (permanent modifier) in combination with Mg(NO3)2 (as aqueous modifier) provided well-defined signal profiles for Cr and Ni determination in coal fly ash samples. A systematic study focused on the use of alternative (less sensitive) lines for obtaining sufficient sensitivity. The following analytical lines were used: Bi 306.8 nm, Cd 326.1 nm, Cr 520.6 nm, Ni 305.1 nm, Pb 205.3 nm, Sb 206.8 nm, Sn 300.9 nm, Zn 307.6 nm. The limits of detection were 0.057 μg g− 1 for Bi, 0.21 μg g− 1 for Cd, 1.1 μg g− 1 for Cr, 1.4 μg g− 1 for Ni, 4.0 μg g− 1 for Pb, 0.13 μg g− 1 for Sb, 0.33 μg g− 1 for Sn and 16 μg g− 1 for Zn, respectively.
Keywords :
Chemical modifiers , Analytical lines , Direct solid sampling , ET AAS , Brown coal fly ash
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy
Serial Year :
2012
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy
Record number :
1684090
Link To Document :
بازگشت