• Title of article

    Multipumping flow system for improving hydride generation atomic fluorescence spectrometric determinations

  • Author/Authors

    Lَpez-Garcيa، نويسنده , , Ignacio and Ruيz-Alcaraz، نويسنده , , Irene and Hernلndez-Cَrdoba، نويسنده , , Manuel، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2006
  • Pages
    5
  • From page
    368
  • To page
    372
  • Abstract
    The advantages of using membrane micropumps rather than peristaltic pumps to introduce both sample and reagent solutions for hydride generation atomic fluorescence spectrometry are discussed. Arsenic was used as a test analyte to check the performance of the proposed manifold. Sample and reagent consumption was reduced 8–9 fold compared with continuous mode measurements made with peristaltic pumps, with no deterioration in sensitivity. The calibration graph was linear in the 0.05 to 2.5 μg l− 1 As range using peak area as the analytical signal and maximum gain in the detector setting. A limit of detection (3σ) of 0.02 μg l− 1 and relative standard deviation values close to 2% for 10 independent measurements of a 1 μg l− 1 As solution were obtained. The sampling frequency increased from 45 to 102 h− 1 with the subsequent saving in carrier gas used and reduction in wastes generated. The instrumental modification, which could be used for other elements currently determined by atomic fluorescence spectrometry, will permit hydride generators of more reduced dimensions to be constructed.
  • Keywords
    Multipumping flow system , Arsenic , Hydride generation , AFSatomic fluorescence spectrometry , instrumentation
  • Journal title
    Spectrochimica Acta Part B Atomic Spectroscopy
  • Serial Year
    2006
  • Journal title
    Spectrochimica Acta Part B Atomic Spectroscopy
  • Record number

    1686669