Title of article :
Determination of Te in soldering tin using continuous flowing electrochemical hydride generation atomic fluorescence spectrometry
Author/Authors :
Jiang، نويسنده , , Xianjuan and Gan، نويسنده , , Wuer and Han، نويسنده , , Suping and He، نويسنده , , Youzhao He، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Pages :
4
From page :
710
To page :
713
Abstract :
An electrochemical hydride generation system was developed for the detection of Te by coupling an electrochemical hydride generator with atomic fluorescence spectrometry. Since TeH2 is unstable and easily decomposes in solution, a reticular W filament cathode was used in the present system. The TeH2 generated on the cathode surface was effectively driven out by sweeping gas from the cathode chamber. In addition, a low temperature electrochemical cell (10 °C) was applied to reduce the decomposition of TeH2 in solution. The limit of detection (LOD) was 2.2 ng ml− 1 and the relative standard deviation (RSD) was 3.9% for nine consecutive measurements of standard solution. This method was successfully employed for determination of Te in soldering tin material.
Keywords :
Atomic fluorescence spectroscopy , TE , Electrochemical hydride generation
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy
Serial Year :
2008
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy
Record number :
1687312
Link To Document :
بازگشت