• Title of article

    Sensitive determination of bismuth by flame atomic absorption spectrometry using atom trapping in a slotted quartz tube and revolatilization with organic solvent pulse

  • Author/Authors

    K?l?nç، نويسنده , , Ersin and Bak?rdere، نويسنده , , Sezgin and Ayd?n، نويسنده , , F?rat and Ataman، نويسنده , , O. Yavuz، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2012
  • Pages
    5
  • From page
    84
  • To page
    88
  • Abstract
    Sensitivity of flame atomic absorption spectrometry (FAAS) for Bi determination was improved by slotted quartz tube (SQT) that was used also for atom trapping (AT). The trapped analyte was released by aspirating a small volume of organic solvent after a reasonable analyte collection time. Sensitivity was improved by 2.9 times by SQT-FAAS and 256 times by SQT-AT-FAAS with respect to FAAS. Optimum trapping period was found to be 6.0 min (36.0 mL of solution). Limit of detection (LOD) for SQT-AT-FAAS was found to be 1.6 ng mL− 1. %RSD was calculated as 4.0% for five replicate measurements of 7.5 ng mL− 1 Bi by SQT-AT-FAAS. Accuracy of the method developed was checked by analyzing a standard reference material of simulated fresh water (NIST 1643e) and result found was in good agreement with the certified one. The method can be applied in any laboratory equipped with a flame AA spectrometer. The consumption of time and sample volume is fairly low and application is simple and easy.
  • Keywords
    Bismuth , Slotted quartz tube , atom trapping , Flame AAS
  • Journal title
    Spectrochimica Acta Part B Atomic Spectroscopy
  • Serial Year
    2012
  • Journal title
    Spectrochimica Acta Part B Atomic Spectroscopy
  • Record number

    1688509