Title of article :
Confocal X-ray fluorescence micro-spectroscopy experiment in tilted geometry
Author/Authors :
Czyzycki، نويسنده , , Mateusz and Wrobel، نويسنده , , Pawel and Lankosz، نويسنده , , Marek، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2014
Abstract :
This paper provides a generalized mathematical model to describe the intensity of primary X-ray fluorescence radiation collected in the tilted confocal geometry mode, where the collimating optics is rotated over an angle relative to a horizontal plane. The influence of newly introduced terms, which take into account the tilted geometry mode, is discussed. The model is verified with a multi-layer test sample scanned in depth. It is proved that for low-Z matrices, the rotation of the detection channel does not induce any significant differences in a reconstruction of the thickness and chemical composition of layers, so that it may safely be ignored.
Keywords :
Confocal X-ray fluorescence spectroscopy , Tilted geometry , Multi-layer materials
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy
Journal title :
Spectrochimica Acta Part B Atomic Spectroscopy