Title of article :
A comparative study of the performance of metallic and FGM tibia tray components in total knee replacement joints
Author/Authors :
Enab، نويسنده , , Tawakol A.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2012
Abstract :
Finite element analysis has been used to predict knee implant biomechanical behavior under various loading conditions. Two-dimensional finite element models have been developed which include the artificial knee and portions of the surrounding biological materials to investigate the required design as a functionally graded material (FGM) tibia tray. The gradation of elastic modulus is changed vertically and horizontally in order to find the optimal gradation direction. The investigation has shown that the use of FGM tibia tray will give excellent performance, and will increase the life of the total knee prosthesis.
Keywords :
Functionally graded material (FGM) , Total knee replacement (TKR) , Tibia tray , Finite element analysis , Biomaterials
Journal title :
Computational Materials Science
Journal title :
Computational Materials Science