Title of article :
Determination of celecoxib in human plasma by normal-phase high-performance liquid chromatography with column switching and ultraviolet absorbance detection
Author/Authors :
Rose، نويسنده , , M.J. and Woolf، نويسنده , , E.J and Matuszewski، نويسنده , , B.K.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Abstract :
A method is described for the determination of celecoxib in human plasma. Samples were extracted using 3M Empore membrane extraction cartridges and separated under normal-phase HPLC conditions using a Nucleosil-NO2 (150×4.6 mm, 5 μm) column. Detection was accomplished using UV absorbance at 260 nm. The HPLC method included a column switching procedure, in which late eluting compounds were diverted to waste, to reduce run-time to 12 min. The assay was linear in the concentration range of 25–2000 ng/ml when 1-ml aliquots of plasma were extracted. Recoveries of celecoxib were greater than 91% over the calibration curve range. Intraday precision and accuracy for this assay were 5.7% C.V. or better and within 2.3% of nominal, respectively. The assay was used to analyze samples collected during human clinical studies.
Journal title :
Journal of Chromatography B Biomedical Sciences and Applications
Journal title :
Journal of Chromatography B Biomedical Sciences and Applications