• Title of article

    Development and validation of a new high-performance liquid chromatographic estimation method of meloxicam in biological samples

  • Author/Authors

    Velpandian، نويسنده , , Thirumurthy and Jaiswal، نويسنده , , Jagdish and Bhardwaj، نويسنده , , Rajinder K. and Gupta، نويسنده , , Suresh K.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2000
  • Pages
    6
  • From page
    431
  • To page
    436
  • Abstract
    A simple, HPLC method was developed to estimate meloxicam (COX-2 inhibitor) using piroxicam as the internal standard. The mobile phase containing methanol, acetonitrile and an aqueous solution of diammonium hydrogenorthophosphate (50 mM) in the ratio of 4:1:5 was pumped at the rate 1 ml/min. Lichrocart RP-18 (125×4 mm) was used as an analytical column and the analytes were detected at 364 nm using a UV detector. Acidified plasma samples were extracted with chloroform, evaporated to dryness, reconstituted in the mobile phase and then a volume of 10 μl of the prepared sample was injected in the column. The retention time of meloxicam and piroxicam was found to be 2.7 and 1.9, respectively. This method showed an accuracy of 102.3% at 0.52 μg/ml and was capable of detecting a minimum concentration of 0.029 μg/ml meloxicam from biological samples. The analytical method was successfully utilized for estimating meloxicam in biological samples.
  • Keywords
    Meloxicam
  • Journal title
    Journal of Chromatography B Biomedical Sciences and Applications
  • Serial Year
    2000
  • Journal title
    Journal of Chromatography B Biomedical Sciences and Applications
  • Record number