• Title of article

    Purge-and-trap gas chromatographic determination of styrene in urine and blood: Application to exposed workers

  • Author/Authors

    Prieto، نويسنده , , M.J and Berenguer، نويسنده , , V and Marhuenda، نويسنده , , D and Cardona، نويسنده , , A، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2000
  • Pages
    6
  • From page
    301
  • To page
    306
  • Abstract
    A simple purge-and-trap gas chromatographic method with flame ionization detection was developed for the determination of styrene in urine and blood. Styrene present in a 5 ml sample at room temperature was swept by helium at 40 ml/min for 11 min, trapped on a Tenax trap, desorbed by heating, cryofocused, and injected by flash heating into a DB-5 capillary GC column. The oven temperature program was from 80°C, held for 8 min, to 120°C at 5°C/min, and then held for 2 min. The detector temperature was 250°C. The calibration curves were linear in the range of 2.5–15 ppb styrene in urine and 25–150 ppb in blood. The detection limits calculated were 0.4 μg/l in urine and 0.6 μg/l in blood. The coefficients of variations within the day and day-to-day were 3 and 3.1%, respectively, for 2.5 ppb of styrene in urine, and 1 and 1.6% for 25 ppb of styrene in blood. The results obtained from samples taken from workers exposed to styrene were reported.
  • Keywords
    styrene
  • Journal title
    Journal of Chromatography B Biomedical Sciences and Applications
  • Serial Year
    2000
  • Journal title
    Journal of Chromatography B Biomedical Sciences and Applications
  • Record number

    1703294