Title of article :
Determination of moxifloxacin (BAY 12-8039) in plasma and lung tissue by high-performance liquid chromatography with ultraviolet detection using a fully automated extraction method with a new polymeric cartridge
Author/Authors :
Lemoine، نويسنده , , T and Breilh، نويسنده , , D and Ducint، نويسنده , , Jean Dubrez، نويسنده , , J and Jougon، نويسنده , , J and Velly، نويسنده , , J.F and Saux، نويسنده , , M.C، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Pages :
8
From page :
247
To page :
254
Abstract :
The aim of this study was to develop a high-performance liquid chromatographic (HPLC) assay for the determination of moxifloxacin in human plasma and lung tissue. The assay was based on HPLC with a Supelcosil ABZ+ column and ultraviolet detection set at a wavelength of 296 nm. The extraction procedure was characterized by a fully automated liquid–solid extraction using an OASIS column for the solid phase. The assay has been found to be linear and validated over the concentration range 3.2 to 0.025 μg/ml for moxifloxacin in plasma and from 16 to 0.25 μg/g for moxifloxacin in lung tissue. In future, the assay will support the pharmacokinetic study of the penetration of moxifloxacin in human lung tissue.
Keywords :
Moxifloxacin
Journal title :
Journal of Chromatography B Biomedical Sciences and Applications
Serial Year :
2000
Journal title :
Journal of Chromatography B Biomedical Sciences and Applications