Title of article
Rapid and sensitive high-performance liquid chromatographic method for busulfan assay in plasma
Author/Authors
Nathalie Bleyzac، نويسنده , , Nathalie and Barou، نويسنده , , Philippe and Aulagner، نويسنده , , Gilles، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2000
Pages
6
From page
427
To page
432
Abstract
A reversed-phase liquid chromatographic method with ultraviolet detection has been developed to determine busulfan concentrations in plasma of children undergoing bone marrow transplantation. Plasma samples (200 μl) containing busulfan and 1,6-bis(methanesulfonyloxy)hexane as an internal standard were prepared by a simple derivatization method with diethyldithiocarbamate followed by extraction with ethyl acetate and solid-phase purification on C8 columns conditioned with methanol and water and eluted with acetonitrile (recovery 99%). Chromatography was accomplished using a Hypersil octadecylsilyl column (10 cm×4.6 mm I.D.) and a mobile phase of acetonitrile, tetrahydrofuran and distilled water (65:5:30, v/v). The limit of detection was 25 ng/ml (signal-to-noise ratio of 5). Calibration curves were linear up to 25 000 ng/ml. Intra-day and inter-day coefficients of variation of the assay were ≤5%. This method was used to analyse busulfan plasma concentrations after oral administration within the framework of therapeutic drug monitoring and pharmacokinetic studies in children.
Keywords
Busulfan
Journal title
Journal of Chromatography B Biomedical Sciences and Applications
Serial Year
2000
Journal title
Journal of Chromatography B Biomedical Sciences and Applications
Record number
1703453
Link To Document