Title of article
Automated solid-phase extraction method for the determination of atovaquone in plasma and whole blood by rapid high-performance liquid chromatography
Author/Authors
Lindegهrdh، نويسنده , , N and Bergqvist، نويسنده , , Y، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2000
Pages
9
From page
9
To page
17
Abstract
A bioanalytical method for the determination of atovaquone in plasma and whole blood by solid-phase extraction and high-performance liquid chromatography has been developed and validated. A structurally similar internal standard was added and protein was precipitated from plasma and whole blood with acetonitrile before being loaded on to a C8 solid-phase extraction column. Atovaquone and internal standard were analysed by high-performance liquid chromatography on a C18 J’Sphere ODS-M80 (150×4.0 mm) column with mobile phase acetonitrile–phosphate buffer, 0.01 M, pH 7.0 (65:35, v/v) and UV detection at 277 nm. The intra-assay precisions for plasma and whole blood were 2.2% and 1.9% respectively at 12 μM and 6.0% and 5.6% respectively at 0.75 μM. The inter-assay precisions for plasma and whole blood were 1.4% and 2.1% respectively at 12 μM and 4.9% and 3.4% respectively at 0.75 μM. The lower limit of quantification in plasma and whole blood were 150 nM. The limit of detection in plasma and whole blood were 30 nM.
Keywords
Atovaquone
Journal title
Journal of Chromatography B Biomedical Sciences and Applications
Serial Year
2000
Journal title
Journal of Chromatography B Biomedical Sciences and Applications
Record number
1703605
Link To Document