• Title of article

    Stereoselective high-performance liquid chromatographic determination of ketamine and its active metabolite, norketamine, in human plasma

  • Author/Authors

    Yanagihara، نويسنده , , Yoshitsugu and Ohtani، نويسنده , , Michiteru and Kariya، نويسنده , , Satoru and Uchino، نويسنده , , Katsuyoshi and Aoyama، نويسنده , , Takao and Yamamura، نويسنده , , Yoshikazu and Iga، نويسنده , , Tatsuji، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2000
  • Pages
    5
  • From page
    227
  • To page
    231
  • Abstract
    A stereoselective high-performance liquid chromatographic method for the determination of the enantiomers of ketamine and its active metabolite, norketamine, in human plasma is described. The compounds were extracted from plasma by liquid–liquid extraction three times in a combination of cyclohexane with 2.5 M NaOH, 1 mM HCl and 1 M carbonate buffer. Stereoselective separation was achieved on a Chiralcel OD column with a mobile phase of n-hexane–2-propanol (98:2, v/v). The detection wavelength was 215 nm. The lower limits of the determination of the method were 5 ng/ml for ketamine and 10 ng/ml for norketamine. The intra- and inter-day coefficients of variation ranged from 2.9 to 9.8% and from 3.4 to 10.7% for all compounds, respectively. The method was sensitive and sufficiently reproducible for stereoselective monitoring of ketamine and norketamine in human plasma during pharmacokinetic studies after the administration of ketamine for analgesia.
  • Journal title
    Journal of Chromatography B Biomedical Sciences and Applications
  • Serial Year
    2000
  • Journal title
    Journal of Chromatography B Biomedical Sciences and Applications
  • Record number

    1703985