• Title of article

    Determination of scopolamine in human serum and microdialysis samples by liquid chromatography–tandem mass spectrometry

  • Author/Authors

    Oertel، نويسنده , , R and Richter، نويسنده , , K and Ebert، نويسنده , , U and Kirch، نويسنده , , W، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2001
  • Pages
    8
  • From page
    121
  • To page
    128
  • Abstract
    A liquid chromatographic-tandem mass spectrometric (LC–MS–MS) method with a rapid and simple sample preparation was developed for the determination of scopolamine in biological fluids. Scopolamine and the internal standard atropine in serum samples were extracted and cleaned up by using an automated solid phase extraction method. Microdialysis samples were directly injected into the LC–MS system. The mass spectrometer was operated in the multi reaction monitoring mode. A good linear response over the range of 20 pg/ml to 5 ng/ml was demonstrated. The accuracy for added scopolamine ranged from 95.0 to 104.0%. The lower limit of quantification was 20 pg/ml. This method is suitable for pharmacokinetic studies.
  • Keywords
    Scopolamine
  • Journal title
    Journal of Chromatography B Biomedical Sciences and Applications
  • Serial Year
    2001
  • Journal title
    Journal of Chromatography B Biomedical Sciences and Applications
  • Record number