• Title of article

    New sensitive assay of vancomycin in human plasma using high-performance liquid chromatography and electrochemical detection

  • Author/Authors

    Favetta، نويسنده , , P. and Guitton، نويسنده , , J. and Bleyzac، نويسنده , , Mohamed N. and Dufresne، نويسنده , , C. and Bureau، نويسنده , , J.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2001
  • Pages
    6
  • From page
    377
  • To page
    382
  • Abstract
    A method using reversed-phase high-performance liquid chromatography with electrochemical detection for the analysis of vancomycin in human plasma was developed. Chromatographic conditions included an octadecyl column, a mobile phase of acetonitrile–sodium phosphate buffer (pH 7) (12:88), a total run time of 12 min, and coulometric electrochemical detection at +700 mV. Linear detector response was found in the range 5–100 μg ml−1 after a 1:80 dilution or from 0.5 to 50 μg ml−1 after a 1:20 dilution of the samples. In both cases the correlation coefficient (r) of the calibration curve standard was better than 0.995. Vancomycin determination was based on a denaturation of plasma proteins with methanol, then a dilution with mobile phase was performed. Recovery of vancomycin from plasma was 103.1±3.9%, and no interference from commonly used drugs or endogenous compounds was observed. A significant correlation was shown with the EMIT assay (r=0.92, P<0.001) using clinical samples from children. This HPLC technique is simple, sensitive, rapid, precise, selective and requires only 100 μl of plasma for completion.
  • Keywords
    Vancomycin
  • Journal title
    Journal of Chromatography B Biomedical Sciences and Applications
  • Serial Year
    2001
  • Journal title
    Journal of Chromatography B Biomedical Sciences and Applications
  • Record number

    1704453