Title of article
Novel liquid chromatographic–tandem mass spectrometric methods using silica columns and aqueous–organic mobile phases for quantitative analysis of polar ionic analytes in biological fluids
Author/Authors
Naidong، نويسنده , , Weng and Shou، نويسنده , , Wilson and Chen، نويسنده , , Yu-Luan and Jiang، نويسنده , , Xiangyu، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2001
Pages
13
From page
387
To page
399
Abstract
Use of silica stationary phase and aqueous–organic mobile phases could significantly enhance LC–MS–MS method sensitivity. The LC conditions were compatible with MS detection. Analytes with basic functional groups were eluted with acidic mobile phases and detected by MS in the positive ion mode. Analytes with acid functional groups were eluted with mobile phases at neutral pH and detected by MS in the negative ion mode. Analytes poorly retained on reversed-phase columns showed good retention on silica columns. Compared with reversed-phase LC–MS–MS, 5–8-fold sensitivity increases were observed for basic polar ionic compounds when using silica columns and aqueous–organic mobile phase. Up to a 20-fold sensitivity increase was observed for acidic polar ionic compounds. Silica columns and aqueous–organic mobile phases were used for assaying nicotine, cotinine, and albuterol in biological fluids.
Keywords
Cotinine , Albuterol , nicotine
Journal title
Journal of Chromatography B Biomedical Sciences and Applications
Serial Year
2001
Journal title
Journal of Chromatography B Biomedical Sciences and Applications
Record number
1704906
Link To Document