• Title of article

    Determination of amlodipine in human plasma by high-performance liquid chromatography with fluorescence detection

  • Author/Authors

    Tatar، نويسنده , , Sevgi and Atmaca، نويسنده , , Sedef، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2001
  • Pages
    6
  • From page
    305
  • To page
    310
  • Abstract
    A sensitive and specific HPLC method has been developed for the assay of amlodipine in human plasma. The assay involves derivatization with 4-chloro-7-nitrobenzofurazan (NBD-Cl), solid-phase extraction on a silica column and isocratic reversed-phase chromatography with fluorescence detection. Nortriptyline hydrochloride was used as an internal standard. The assay was linear over the concentration range of 0.25–18.00 ng/ml. Both of the within-day and day-to-day reproducibility and accuracy were less than 11.80% and 12.00%, respectively. The plasma profile following a single administration of 10 mg amlodipine to a healthy volunteer was presented.
  • Keywords
    Amlodipine , 4-Chloro-7-nitrobenzofurazan
  • Journal title
    Journal of Chromatography B Biomedical Sciences and Applications
  • Serial Year
    2001
  • Journal title
    Journal of Chromatography B Biomedical Sciences and Applications
  • Record number

    1705435