Title of article :
Gas chromatographic–mass spectrometric method for quantitation of phenylalanine and tyrosine in serum
Author/Authors :
Laurens، نويسنده , , J.B and Mbianda، نويسنده , , X.Y and Ubbink، نويسنده , , J.B and Vermaak، نويسنده , , W.J.H، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Pages :
10
From page :
127
To page :
136
Abstract :
A validated gas chromatographic–mass spectrometric method for quantitation of phenylalanine and tyrosine in serum is described. Quantitation of phenylalanine and tyrosine with a non-labelled non-endogenous internal standard, l-2-chlorophenylalanine, compared favourably with isotope dilution mass spectrometric quantitation. The 95% reference ranges for phenylalanine, tyrosine and the phenylalanine–tyrosine molar ratio in neonate cord blood serum were determined by isotope dilution mass spectrometry and were found to be 77.1–144.7, 33.3–109.3 μmol/l and 1.1–3.0, respectively.
Keywords :
Phenylketonuria , phenylalanine , tyrosine
Journal title :
Journal of Chromatography B Biomedical Sciences and Applications
Serial Year :
2001
Journal title :
Journal of Chromatography B Biomedical Sciences and Applications
Record number :
1705818
Link To Document :
بازگشت