Title of article :
A pharmacokinetic interpretation of increasing concentrations of DEHP in haemodialysed patients
Author/Authors :
Dine، نويسنده , , Thierry and Luyckx، نويسنده , , Michel and Gressier، نويسنده , , Bernard and Brunet، نويسنده , , Claude and Souhait، نويسنده , , Jean and Nogarede، نويسنده , , Serge and Vanpoucke، نويسنده , , Jacques and Courbon، نويسنده , , Frédéric and Plusquellec، نويسنده , , Yves and Houin، نويسنده , , Georges، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Pages :
9
From page :
157
To page :
165
Abstract :
The degree of exposure to DEHP was assessed in 11 patients with chronic renal failure undergoing maintenance haemodialysis. The amount of DEHP leached from the dialyser during a 4-h dialysis session was estimated by monitoring the DEHP blood concentration using a HPLC method. When a patient undergoes a dialysis treatment, the concentration of di-2-ethylhexyl phthalate (DEHP) in venous blood is increased when the blood crosses through the dialysis apparatus. This increase may be explained either because DEHP is not extracted by the dialyser or because DEHP comes from the dialysis bath due to contact of blood against plasticized pipes. To explain the increasing concentration of DEHP during treatment of renal failure using plasticized tubing, we propose a pharmacokinetic compartmental model in order to fit raw data obtained from dialysed patients and to get the amount of DEHP which enters the body by AUC calculations. Results obtained after HPLC analysis show a high degree of interpatient variability in DEHP retained. This amount can reach a toxicity level because of repetitive dialysis treatments over prolonged periods of time. In the coming years, it seems necessary to reconsider the use of DEHP as a plasticizer in medical devices. Highly unacceptable amounts of DEHP leached during the dialysis session could be easily avoided by careful selection of haemodialysis tubing.
Keywords :
Haemodialysis , DEHP , Area under curve , Compartmental model , TOXICITY
Journal title :
Medical Engineering and Physics
Serial Year :
2000
Journal title :
Medical Engineering and Physics
Record number :
1727109
Link To Document :
بازگشت