Title of article :
Implementation of non-invasive brain physiological monitoring concepts
Author/Authors :
Ragauskas، نويسنده , , Arminas and Daubaris، نويسنده , , Gediminas and Ragaisis، نويسنده , , Vytautas and Petkus، نويسنده , , Vytautas، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Abstract :
The paper presents innovative methods and technology for non-invasive intracranial hemodynamics monitoring based on the measurement of brain parenchyma acoustic properties. The clinical investigation of new technology shows the similarity between the invasively recorded intracranial pressure (ICP) and non-invasively recorded intracranial blood volume (IBV) pulse waves, slow waves and slow trends under intensive care unit (ICU) conditions. Also, the applicability of the non-invasive IBV slow wave monitoring technique for cerebrovascular autoregulation non-invasive long-term monitoring is demonstrated by theoretical and experimental studies.
Keywords :
Non-invasive monitoring , Cerebrovascular autoregulation , Intracranial pressure , Ultrasonic simulation , Time-of-flight method
Journal title :
Medical Engineering and Physics
Journal title :