Title of article :
A novel model of the pulse decay method for measurement of local tissue blood perfusion
Author/Authors :
Yang، نويسنده , , Kun and Liu، نويسنده , , Wei، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Pages :
9
From page :
215
To page :
223
Abstract :
The blood perfusion of biological tissue is a basic parameter of physiology and medical engineering. A novel average temperature model (ATM) is presented in this paper to measure the blood perfusion of tissue based on the thermal pulse-decay method. Differing from the existing point source model (PSM) and spherical source model (SSM), the probe bead average temperature analytical solution is derived and used to estimate the blood perfusion. The blood perfusion prediction errors caused by the approximate assumptions used in each model are studied using the numerical experiment method. Contributions of the tissue parameters, probe thermistor bead parameters, and measurement parameters, such as tissue thermal conductivity, tissue thermal diffusivity, blood perfusion rate, bead thermal conductivity, bead radius, measurement time, and thermal pulse length are discussed. The predicting accuracy of the ATM model is compared with the PSM model and SSM model. The results show that, for all the cases tested, the ATM model is better than the other two.
Keywords :
Blood perfusion , Average temperature model , Thermal pulse decay method , prediction error , Numerical experiment
Journal title :
Medical Engineering and Physics
Serial Year :
2004
Journal title :
Medical Engineering and Physics
Record number :
1728236