• Title of article

    The effect of duty cycle and frequency on muscle torque production using kilohertz frequency range alternating current

  • Author/Authors

    Ward، نويسنده , , Alex R and Robertson، نويسنده , , Valma J and Ioannou، نويسنده , , Harry، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    11
  • From page
    569
  • To page
    579
  • Abstract
    We investigated the frequency and duty cycle dependence of maximal electrically induced torque (MEIT) of the wrist extensors. Fifty hertz burst modulated sinusoidal alternating current (AC) in the frequency range 0.5–20 kHz was used, with duty cycles ranging from a minimum (one cycle) to maximum (continuous AC). were similar at low frequencies but decreased markedly above 2.5 kHz. MEITs also decreased markedly above a 20% duty cycle. Subjective reports of discomfort were fewest at 4 kHz and at duty cycles in the range 20–25%. nclusion is that for maximum torque production, a frequency of 1 kHz and a duty cycle of 20% are indicated. When comfort is a major consideration, a frequency of 2.5 kHz provides an acceptable trade-off between MEIT and comfort. ndings also suggest that low duty cycle, burst modulated AC stimulation may be more effective than stimulation using conventional low-frequency pulsed current.
  • Keywords
    torque , Electrical stimulation , Alternating current , Duty cycle , Frequency
  • Journal title
    Medical Engineering and Physics
  • Serial Year
    2004
  • Journal title
    Medical Engineering and Physics
  • Record number