Title of article :
A system for the delivery of programmable, adaptive stimulation intensity envelopes for drop foot correction applications
Author/Authors :
Breen، نويسنده , , P.P. and O’Keeffe، نويسنده , , D.T. and Conway، نويسنده , , R. and Lyons، نويسنده , , G.M.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Pages :
10
From page :
177
To page :
186
Abstract :
We describe the design of an intelligent drop foot stimulator unit for use in conjunction with a commercial neuromuscular electrical nerve stimulation (NMES) unit, the NT2000. The developed micro-controller unit interfaces to a personal computer (PC) and a graphical user interface (GUI) allows the clinician to graphically specify the shape of the stimulation intensity envelope required for a subject undergoing drop foot correction. The developed unit is based on the ADuC812S micro-controller evaluation board from Analog Devices and uses two force sensitive resistor (FSR) based foot-switches to control application of stimulus. The unit has the ability to display to the clinician how the stimulus intensity envelope is being delivered during walking using a data capture capability. The developed system has a built-in algorithm to dynamically adjust the delivery of stimulus to reflect changes both within the gait cycle and from cycle to cycle. Thus, adaptive control of stimulus intensity is achieved.
Keywords :
Stimulation intensity envelope , Drop foot stimulator , Functional electrical stimulation (FES)