Title of article :
Gradient coil design using Bi-2223 high temperature superconducting tape for magnetic resonance imaging
Author/Authors :
Yuan، نويسنده , , Jing and Shen، نويسنده , , G.X.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Pages :
7
From page :
442
To page :
448
Abstract :
Bi-2223 tape is a low-cost, long-length high temperature superconducting material. The new application of Bi-2223 tape for gradient coils for magnetic resonance imaging (MRI) is studied in this paper. Because Bi-2223 tape has a much higher critical current density and a much lower power loss than copper, a Bi-2223 high temperature superconductor (HTS) gradient coil shows great advantages in high gradient strengths, long continuous gradient rating, and free of cooling system over a copper gradient coil. The power losses of Bi-2223 tapes at working frequencies of gradient coils are compared to copper. The critical current degradation of tapes is also discussed. A prototype of HTS tape gradient coil is fabricated and its gradient field distributions are measured. Technical issues of resistance, gradient strength, continuous gradient rating, and cryostat are also discussed.
Keywords :
Bi-2223 tape , critical current , Power loss , gradient coil , MAGNETIC RESONANCE IMAGING
Journal title :
Medical Engineering and Physics
Serial Year :
2007
Journal title :
Medical Engineering and Physics
Record number :
1729416
Link To Document :
بازگشت