Title of article :
Ultrasound low-velocity flow estimations using cross-correlation and decorrelation: A thread phantom study
Author/Authors :
Pan، نويسنده , , Ton-Tai and Chiang، نويسنده , , Huihua Kenny Chiang، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Abstract :
In this paper, two time-domain flow velocity measurement methods, cross-correlation and decorrelation, were implemented and compared in measuring low blood-flow velocities, ranging from 0.05 to 5 cm/s, at known beam-vessel angles of 50–90° in a thread phantom. Both methods achieved a significant correlation between the measured and the known velocities (r = 0.99; P < 0.001) in the thread phantom measurements, at 50–70° beam-vessel angles. Although the cross-correlation method showed a significant correlation (r > 0.99; P < 0.001), this method caused large errors (>45%) at 80° and 90°, whereas the decorrelation method not only achieved a significant correlation (r > 0.99; P < 0.001) but also resulted in smaller errors (<19%) at 80° and 90°. This study demonstrates the potential of time-domain decorrelation method for measuring low flow velocity at large beam-vessel angles near perpendicular conditions.
Keywords :
Beam-vessel angle , Low flow velocity , Time-Domain , cross-correlation , Decorrelation
Journal title :
Medical Engineering and Physics
Journal title :
Medical Engineering and Physics