Title of article
Development of a synthetic phantom for the selection of optimal scanning parameters in CAD–CT colonography
Author/Authors
Chowdhury، نويسنده , , Tarik A. and Whelan، نويسنده , , Paul F. and Ghita، نويسنده , , Ovidiu and Sezille، نويسنده , , Nicolas and Foley، نويسنده , , Shane، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2007
Pages
10
From page
858
To page
867
Abstract
The aim of this paper is to present the development of a synthetic phantom that can be used for the selection of optimal scanning parameters in computed tomography (CT) colonography. In this paper we attempt to evaluate the influence of the main scanning parameters including slice thickness, reconstruction interval, field of view, table speed and radiation dose on the overall performance of a computer aided detection (CAD)–CTC system. From these parameters the radiation dose received a special attention, as the major problem associated with CTC is the patient exposure to significant levels of ionising radiation. To examine the influence of the scanning parameters we performed 51 CT scans where the spread of scanning parameters was divided into seven different protocols. A large number of experimental tests were performed and the results analysed. The results show that automatic polyp detection is feasible even in cases when the CAD–CTC system was applied to low dose CT data acquired with the following protocol: 13 mAs/rotation with collimation of 1.5 mm × 16 mm, slice thickness of 3.0 mm, reconstruction interval of 1.5 mm, table speed of 30 mm per rotation. The CT phantom data acquired using this protocol was analysed by an automated CAD–CTC system and the experimental results indicate that our system identified all clinically significant polyps (i.e. larger than 5 mm).
Keywords
radiation dose , Synthetic phantom , Polyp detection , CT colonography
Journal title
Medical Engineering and Physics
Serial Year
2007
Journal title
Medical Engineering and Physics
Record number
1729567
Link To Document