Title of article
Use of multiple wearable inertial sensors in upper limb motion tracking
Author/Authors
Zhou، نويسنده , , Huiyu and Stone، نويسنده , , Thomas and Hu، نويسنده , , Huosheng and Harris، نويسنده , , Nigel، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2008
Pages
11
From page
123
To page
133
Abstract
This paper presents a new human motion tracking system using two wearable inertial sensors that are placed near the wrist and elbow joints of the upper limb. Each inertial sensor consists of a tri-axial accelerometer, a tri-axial gyroscope and a tri-axial magnetometer. The turning rates of the gyroscope were utilised for localising the wrist and elbow joints on the assumption that the two upper limb segment lengths are known a priori. To determine the translation and rotation of the shoulder joint, an equality-constrained optimisation technique is adopted to find an optimal solution, incorporating measurements from the tri-axial accelerometer and gyroscope. Experimental results demonstrate that this new system, compared to an optical motion tracker, has RMS position errors that are normally less than 0.01 m, and RMS angle errors that are 2.5–4.8°.
Keywords
Inertial sensor , Motion tracking , Upper Limb , Rehabilitation , Optimisation
Journal title
Medical Engineering and Physics
Serial Year
2008
Journal title
Medical Engineering and Physics
Record number
1729739
Link To Document