Title of article
Tribology and total hip joint replacement: Current concepts in mechanical simulation
Author/Authors
Affatato، نويسنده , , Emmanuelle S. and Spinelli، نويسنده , , M. and Zavalloni، نويسنده , , M. and Mazzega-Fabbro، نويسنده , , C. and Viceconti، نويسنده , , M.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2008
Pages
13
From page
1305
To page
1317
Abstract
Interest in the rheology and effects of interacting surfaces is as ancient as man. This subject can be represented by a recently coined word: tribology. This term is derived from the Greek word “tribos” and means the “science of rubbing”. Friction, lubrication, and wear mechanism in the common English language means the precise field of interest of tribology.
f total hip prosthesis is a significant clinical problem that involves, nowadays, a too high a number of patients. In order to acquire further knowledge on the tribological phenomena that involve hip prosthesis wear tests are conducted on employed materials to extend lifetime of orthopaedic implants.
st basic type of test device is the material wear machine, however, a more advanced one may more accurately reproduce some of the in vivo conditions. Typically, these apparatus are called simulators, and, while there is no absolute definition of a joint simulator, its description as a mechanical rig used to test a joint replacement, under conditions approximating those occurring in the human body, is acceptable. Simulator tests, moreover, can be used to conduct accelerated protocols that replicate/simulate particularly extreme conditions, thus establishing the limits of performance for the material. Simulators vary in their level of sophistication and the international literature reveals many interpretations of the design of machines used for joint replacement testing.
aper aims to review the current state of the art of the hip joint simulators worldwide. This is specified through a schematic overview by describing, in particular, constructive solutions adopted to reproduce in vivo conditions.
austive commentary on the evolution and actually existing simulation standards is proposed by the authors. The need of a shared protocol among research laboratories all over the world could lead to a consensus conference.
Keywords
Wear evaluation , Machines for wear tests , Hip simulator studies , Hip implants , Simulator review
Journal title
Medical Engineering and Physics
Serial Year
2008
Journal title
Medical Engineering and Physics
Record number
1730292
Link To Document