Title of article :
Validation of a passive stereophotogrammetry system for imaging of the breast: A geometric analysis
Author/Authors :
Catherwood، نويسنده , , T. and McCaughan، نويسنده , , E. and Greer، نويسنده , , J. E. H. Spence، نويسنده , , R.A.J. and McIntosh، نويسنده , , S.A. and Winder، نويسنده , , R.J.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2011
Abstract :
The overall aim of this study was to assess the accuracy, reproducibility and stability of a high resolution passive stereophotogrammetry system to image a female mannequin torso, to validate measurements made on the textured virtual surface compared with those obtained using manual techniques and to develop an approach to make objective measurements of the female breast. 3D surface imaging was carried out on a textured female torso and measurements made in accordance with the system of mammometrics. Linear errors in measurements were less than 0.5 mm, system calibration produced errors of less than 1.0 mm over 94% over the surface and intra-rater reliability measured by ICC = 0.999. The mean difference between manual and digital curved surface distances was 1.36 mm with maximum and minimum differences of 3.15 mm and 0.02 mm, respectively. The stereophotogrammetry system has been demonstrated to perform accurately and reliably with specific reference to breast assessment.
Keywords :
Validation , Torso , breast reconstruction , STEREOPHOTOGRAMMETRY , 3D
Journal title :
Medical Engineering and Physics
Journal title :
Medical Engineering and Physics