Title of article
A method for characterizing stapes prostheses by their mechanical transfer function
Author/Authors
Sutor، نويسنده , , Alexander and Hornung، نويسنده , , Joachim and Gossler، نويسنده , , Julian and Doellinger، نويسنده , , Michael and Lerch، نويسنده , , Reinhard، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2012
Pages
5
From page
659
To page
663
Abstract
In this contribution, we present and evaluate a method for characterizing stapes prostheses by their mechanical transfer function. The measurements were carried out after a stapedotomy surgery was performed in three human temporal bones conserved in 4% formaldehyde. The inner ear was drained of fluid. Successively, one of three different stapes prostheses was inserted. After such preparation, the prosthesis piston movement compared to the incus movement is measured with laser vibrometry. The magnitude transfer function considered is defined as the amplitude of the prosthesis piston movement compared to the amplitude of the incus movement. Measurements were made in a frequency range from 500 Hz to 4 kHz. The measured amplitudes roughly ranged between 10 nm and 100 nm. A great advantage of the presented method is the fact that only a small portion of the ossicular chain influences the measurement, mainly the joint between the prosthesis and the incus. Furthermore, the usage of cadaver temporal bones allows for an automated measurement setup, long term experiments and the access of measurement positions inapproachable during in vivo measurements. With this method, the different kinds of prostheses could be evaluated on incuses of different diameters.
Keywords
Crimping , Stapes prosthesis , Otosclerosis , Laser vibrometer , Mechanical transfer function
Journal title
Medical Engineering and Physics
Serial Year
2012
Journal title
Medical Engineering and Physics
Record number
1731668
Link To Document