Title of article
Screw locking elements: A means to modify the flexibility of osteoporotic fracture fixation with DCPs without compromising system strength or stability
Author/Authors
Yلnez، نويسنده , , A. and Cuadrado، نويسنده , , Vladimir A. and Carta، نويسنده , , J.A. and Garcés، نويسنده , , G.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2012
Pages
8
From page
717
To page
724
Abstract
This paper analyses whether it is possible to use dynamic compression plates (DCPs) and screw locking elements (SLEs) to vary the flexibility of osteoporotic fracture fixation without compromising the strength and stability of the construct.
ssion, torsion and four-point bending static strength tests were conducted. Cyclic load tests of up to 10,000 load cycles were also carried out to determine stiffness performance. Four fixation systems were mounted onto polyurethane bone models. Group 1 consists of the DCP and six cortical screws. Group 2, idem, but with the addition of two SLEs. Group 3, idem, but with the addition of six SLEs. Group 4 used the locking compression plate (LCP) and locking screws.
sults indicated no significant difference (p > 0.05) in the strength of groups 2–4. It was also observed that the torsional stiffness of group 3 (0.30 N m/°) was higher than that of group 2 (0.23 N m/°) and similar to that of group 4 (0.28 N m/°). Compression stiffness of group 4 (124 N/mm) was higher than that of group 2 (102 N/mm), but lower than that of group 3 (150 N/mm). No notable differences were observed for structural bending stiffness.
concluded that by using the DCP with SLEs it is possible to modify the stiffness of the fixation construct for the repair of osteoporotic fractures and, in this way, facilitate the conditions suitable on secondary bone healing.
Keywords
Osteoporosis , Internal fixation , PEEK , Screw locking elements , Stiffness
Journal title
Medical Engineering and Physics
Serial Year
2012
Journal title
Medical Engineering and Physics
Record number
1731682
Link To Document