• Title of article

    Importance of material model in wall stress prediction in abdominal aortic aneurysms

  • Author/Authors

    Polzer، نويسنده , , Stanislav and Christian Gasser، نويسنده , , T. and Bursa، نويسنده , , Jiri and Staffa، نويسنده , , Robert and Vlachovsky، نويسنده , , Robert de Man، نويسنده , , Vojtech and Skacel، نويسنده , , Pavel، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2013
  • Pages
    8
  • From page
    1282
  • To page
    1289
  • Abstract
    Background s of biomechanical simulation of the abdominal aortic aneurysm (AAA) depend on the constitutive description of the wall. Based on in vitro and in vivo experimental data several constitutive models for the AAA wall have been proposed in the literature. Those models differ strongly from each other and their impact on the computed stress in biomechanical simulation is not clearly understood. s element (FE) models of AAAs from 7 patients who underwent elective surgical repair were used to compute wall stresses. AAA geometry was reconstructed from CT angiography (CT-A) data and patient-specific (PS) constitutive descriptions of the wall were derived from planar biaxial testing of anterior wall tissue samples. In total 28 FE models were used, where the wall was described by either patient-specific or previously reported study-average properties. This data was derived from either uniaxial or biaxial in vitro testing. Computed wall stress fields were compared on node-by-node basis. s ent constitutive models for the AAA wall cause significantly different predictions of wall stress. While study-average data from biaxial testing gives globally the same stress field as the patient-specific wall properties, the material model based on uniaxial test data overestimates the wall stress on average by 30 kPa or about 67% of the mean stress. A quasi-linear description based on the in vivo measured distensibility of the AAA wall leads to a completely altered stress field and overestimates the wall stress by about 75 kPa or about 167% of the mean stress. sion esent study demonstrated that the constitutive description of the wall is crucial for AAA wall stress prediction. Consequently, results obtained using different models should not be mutually compared unless different stress gradients across the wall are not taken into account. Highly nonlinear material models should be preferred when the response of AAA to increased blood pressure is investigated, while the quasi-linear model with high initial stiffness produces negligible stress gradients across the wall and thus, it is more appropriate when response to mean blood pressure is calculated.
  • Keywords
    material model , Pre-stressing , wall stress , FE analysis , Abdominal aneurysm
  • Journal title
    Medical Engineering and Physics
  • Serial Year
    2013
  • Journal title
    Medical Engineering and Physics
  • Record number