Title of article :
Marker-based validation of a biplane fluoroscopy system for quantifying foot kinematics
Author/Authors :
Iaquinto، نويسنده , , Joseph M. and Tsai، نويسنده , , Richard and Haynor، نويسنده , , David R. and Fassbind، نويسنده , , Michael J. and Sangeorzan، نويسنده , , Bruce J. and Ledoux، نويسنده , , William R.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2014
Pages :
6
From page :
391
To page :
396
Abstract :
AbstractIntroduction tereometric analysis has demonstrated its capacity to track precise motion of the bones within a subject during motion. Existing devices for imaging the body in two planes are often custom built systems; we present here the design and marker-based validation of a system that has been optimized to image the foot during gait. s ical modifications were made to paired BV Pulsera C-arms (Philips Medical Systems) to allow unfettered gait through the imaging area. Image quality improvements were obtained with high speed cameras and the correction of image distorting artifacts. To assess the systemʹs accuracy, we placed beads at known locations throughout the imaging field, and used post processing software to calculate their apparent locations. s tion correction reduced overall RMS error from 6.56 mm to 0.17 mm. When tracking beads in static images a translational accuracy of 0.094 ± 0.081 mm and rotational accuracy of 0.083 ± 0.068° was determined. In dynamic trials simulating speeds seen during walking, accuracy was 0.126 ± 0.122 mm. sion curacies and precisions found are within the reported ranges from other such systems. With the completion of marker-based validation, we look to model-based validation of the foot during gait.
Keywords :
Biplane fluoroscopy , X-Ray , Foot , Biomechanics
Journal title :
Medical Engineering and Physics
Serial Year :
2014
Journal title :
Medical Engineering and Physics
Record number :
1732514
Link To Document :
بازگشت