    Soft wearable contact lens sensor for continuous intraocular pressure monitoring

  • Author/Authors

    Chen، نويسنده , , Guo-Zhen and Chan، نويسنده , , Ion-Seng and Leung، نويسنده , , Leo K.K. and Lam، نويسنده , , David C.C.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2014
  • Pages
    6
  • From page
    1134
  • To page
    1139
  • Abstract
    Intraocular pressure (IOP) is a primary indicator of glaucoma, but measurements from a single visit to the clinic miss the peak IOP that may occur at night during sleep. A soft chipless contact lens sensor that allows the IOP to be monitored throughout the day and at night is developed in this study. A resonance circuit composed of a thin film capacitor coupled with a sensing coil that can sense corneal curvature deformation is designed, fabricated and embedded into a soft contact lens. The resonance frequency of the sensor is designed to vary with the lens curvature as it changes with the IOP. The frequency responses and the ability of the sensor to track IOP cycles were tested using a silicone rubber model eye. The results showed that the sensor has excellent linearity with a frequency response of ∼8 kHz/mmHg, and the sensor can accurately track fluctuating IOP. These results showed that the chipless contact lens sensor can potentially be used to monitor IOP to improve diagnosis accuracy and treatment of glaucoma.
  • Keywords
    Intraocular pressure , Glaucoma , Contact lens sensor , Resonance circuit
  • Journal title
    Medical Engineering and Physics
  • Serial Year
    2014
  • Journal title
    Medical Engineering and Physics
  • Record number