    Siegel، نويسنده , , Robert J. and Biner، نويسنده , , Simon and Rafique، نويسنده , , Asim M. and Rinaldi، نويسنده , , Michael and Lim، نويسنده , , Scott and Fail، نويسنده , , Peter and Hermiller، نويسنده , , James and Smalling، نويسنده , , Richard and Whitlow، نويسنده , , Patrick L. and Herrmann، نويسنده , , Howard C. and Foster، نويسنده , , Elyse and Feldman، نويسنده , , Ted and Glower، نويسنده , , Donald an، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2011
  • Pages
    8
  • From page
    1658
  • To page
    1665
  • Abstract
    Objectives jective of this study was to evaluate the acute hemodynamic consequences of mitral valve (MV) repair with the MitraClip device (Abbott Vascular, Menlo Park, California). ound r surgical correction of mitral regurgitation (MR) results in a low cardiac output (CO) state because of an acute increase in afterload remains controversial. The acute hemodynamic consequences of MR reduction with the MitraClip device have not been studied. s luated 107 patients with cardiac catheterization before and immediately following percutaneous MV repair with the MitraClip device. In addition, pre- and post-procedural hemodynamic parameters were studied by transthoracic echocardiography. s lip treatment was attempted in 107 patients, and in 96 (90%) patients, a MitraClip was deployed. Successful MitraClip treatment resulted in: 1) an increase in CO from 5.0 ± 2.0 l/min to 5.7 ± 1.9 l/min (p = 0.003); 2) an increase in forward stroke volume (FSV) from 57 ± 17 ml to 65 ± 18 ml (p < 0.001); and 3) a decrease in systemic vascular resistance from 1,226 ± 481 dyn·s/cm5 to 1,004 ± 442 dyn·s/cm5 (p < 0.001). In addition, there was left ventricular (LV) unloading manifested by a decrease in LV end-diastolic pressure from 11.4 ± 9.0 mm Hg to 8.8 ± 5.8 mm Hg (p = 0.016) and a decrease in LV end-diastolic volume from 172 ± 37 ml to 158 ± 38 ml (p < 0.001). None of the patients developed acute post-procedural low CO state. sions sful MV repair with the MitraClip system results in an immediate and significant improvement in FSV, CO, and LV loading conditions. There was no evidence of a low CO state following MitraClip treatment for MR. These favorable hemodynamic effects with the MitraClip appear to reduce the risk of developing a low CO state, a complication occasionally observed after surgical MV repair for severe MR. (A Study of the Evalve Cardiovascular Valve Repair System Endovascular Valve Edge-to-Edge Repair Study [EVEREST I]; NCT00209339 and EVEREST II; NCT00209274)
  • Keywords
    cardiac output , MitraClip , Hemodynamics , Mitral regurgitation
  • Journal title
    JACC (Journal of the American College of Cardiology)
  • Serial Year
    2011
  • Journal title
    JACC (Journal of the American College of Cardiology)
  • Record number

    1751919