Title of article :
6-Month Clinical Outcomes Following Implantation of the Bioresorbable Everolimus-Eluting Vascular Scaffold in Vessels Smaller or Larger Than 2.5 mm
Author/Authors :
Diletti، نويسنده , , Roberto and Onuma، نويسنده , , Yoshinobu and Farooq، نويسنده , , Vasim and Gomez-Lara، نويسنده , , Josep and Brugaletta، نويسنده , , Salvatore and van Geuns، نويسنده , , Robert Jan and Regar، نويسنده , , Evelyn and de Bruyne، نويسنده , , Bernard and Dudek، نويسنده , , Dariusz and Thuesen، نويسنده , , Leif and Chevalier، نويسنده , , Bernard and McClean، نويسنده , , Dougal and ، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2011
Abstract :
Objectives
estigated the 6-month clinical outcomes after implantation of second-generation 3.0-mm bioresorbable everolimus-eluting vascular scaffolds (BVS) in small coronary vessels (<2.5 mm).
ound
e a novel approach to treating coronary lesions and are untested in small vessels.
s
SORB Cohort B Trial is a multicenter, single-arm, prospective, open-label trial assessing the performance of the second-generation BVS, in which 101 patients were enrolled. The pre-procedural reference vessel diameter (RVD) was assessed by quantitative coronary angiography during post hoc analysis. The vessel size was overestimated, by visual assessment, in 41 patients before implantation of 3.0-mm BVS in vessels with a pre-procedural RVD <2.5 mm. The study population was divided into 2 groups, group I (n = 41) with RVD <2.5 mm and group II (n = 60) with RVD ≥2.5 mm. The composite endpoint of ischemia-driven major adverse cardiac events, defined as ischemia-driven target lesion revascularization, myocardial infarction, or cardiac death, was assessed. Of the 45 patients scheduled for 6-month coronary angiography, 42 patients had the procedure performed, with intravascular ultrasound undertaken in 40 of these patients.
s
onths, no significant differences in ischemia-driven major adverse cardiac events (3 of 41 [7.3%] cases vs. 2 of 60 [3.3%] cases; p = 0.3933) were observed in the small- and large-vessel groups, respectively. No cardiac deaths or episodes of in-scaffold thromboses were seen. Angiographic and intravascular ultrasound follow-up demonstrated no differences in late lumen loss (0.16 ± 0.18 mm vs. 0.21 ± 0.17 mm; p = 0.3525) or percentage lumen area stenosis (17.6 ± 6.0% vs. 19.8 ± 8.5%; p = 0.3643).
sions
cond-generation 3.0-mm BVS appears to be safe in small vessels, with similar clinical and angiographic outcomes compared with those of large vessels.
Keywords :
Outcome , bioresorbable everolimus-eluting vascular scaffold , small vessels
Journal title :
JACC (Journal of the American College of Cardiology)
Journal title :
JACC (Journal of the American College of Cardiology)