Abstract :
Objectives
jective of this study was to compare the cerebral embolic load of filter-protected versus proximal balloon–protected carotid artery stenting (CAS).
ound
ized trials comparing filter-protected CAS with carotid endarterectomy revealed a higher periprocedural stroke rate after CAS. Proximal balloon occlusion may be more effective in preventing cerebral embolization during CAS than filters.
s
ts undergoing CAS with cerebral embolic protection for internal carotid artery stenosis were randomly assigned to proximal balloon occlusion or filter protection. The primary endpoint was the incidence of new cerebral ischemic lesions assessed by diffusion-weighted magnetic resonance imaging. Secondary endpoints were the number and volume of new ischemic lesions and major adverse cardiovascular and cerebral events (MACCE).
s
two consecutive patients (mean age: 71.7 years, 76.4% male) were randomized. Compared with filter protection (n = 31), proximal balloon occlusion (n = 31) resulted in a significant reduction in the incidence of new cerebral ischemic lesions (45.2% vs. 87.1%, p = 0.001). The number (median [range]: 2 [0 to 13] vs. 0 [0 to 4], p = 0.0001) and the volume (0.47 [0 to 2.4] cm3 vs. 0 [0 to 0.84] cm3, p = 0.0001) of new cerebral ischemic lesions were significantly reduced by proximal balloon occlusion. Lesions in the contralateral hemisphere were found in 29.0% and 6.5% of patients (filter vs. balloon occlusion, respectively, p = 0.047). The 30-day MACCE rate was 3.2% and 0% for filter versus balloon occlusion, respectively (p = NS).
sions
s randomized trial of patients undergoing CAS, proximal balloon occlusion as compared with filter protection significantly reduced the embolic load to the brain.
Keywords :
Diffusion-weighted magnetic resonance imaging , cerebral embolic protection , Carotid artery stenting