Title of article :
Does Rubidium-82 PET Have Superior Accuracy to SPECT Perfusion Imaging for the Diagnosis of Obstructive Coronary Disease?: A Systematic Review and Meta-Analysis
Author/Authors :
Mc Ardle، نويسنده , , Brian A. and Dowsley، نويسنده , , Taylor F. and deKemp، نويسنده , , Robert A. and Wells، نويسنده , , George A. and Beanlands، نويسنده , , Rob S.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2012
Abstract :
Objectives
rpose of this study was to evaluate the accuracy of rubidium (Rb)-82 positron emission tomography (PET) for the diagnosis of obstructive coronary artery disease (CAD) in comparison to single-photon emission tomography (SPECT).
ound
dial perfusion imaging is widely used in the assessment of patients with known or suspected CAD. PET using Rb-82 has potential advantages over SPECT that may make it more accurate and that reduce radiation exposure compared with SPECT but has increased costs. Comparisons of these technologies are highly relevant for policy makers and practice guidelines. However, studies directly comparing Rb-82 PET with contemporary SPECT have been limited.
thors therefore undertook a systematic review of studies where either Rb-82 PET or technetium-99m SPECT with both attenuation correction and electrocardiography-gating were used as a diagnostic test for obstructive CAD with invasive coronary angiogram as a reference standard. These technologies were then compared.
s
n PET and 8 SPECT studies (1,344 and 1,755 patients, respectively) met inclusion criteria and pooled accuracy using weighted averages according to the size of the patient population was determined for PET and SPECT with sensitivities of 90% (confidence interval [CI]: 0.88 to 0.92) and 85% (CI: 0.82 to 0.87) and specificities of 88% (CI: 0.85 to 0.91) and 85% (CI: 0.82 to 0.87), respectively. Summary receiver-operating characteristic curves were computed: area under the curve was 0.95 and 0.90 for PET and SPECT, respectively (p < 0.0001). There was heterogeneity among study populations and some studies were limited by referral bias.
sions
PET is accurate for the detection of obstructive CAD and, despite advances in SPECT technology, remains superior. More widespread use of Rb-82 PET may be beneficial to improve CAD detection.
Keywords :
CAD , PET , accuracy , SPECT , diagnosis
Journal title :
JACC (Journal of the American College of Cardiology)
Journal title :
JACC (Journal of the American College of Cardiology)