Abstract :
The use of an artificial joint is always related to a certain amount of wear. Its biological effects, e.g., the osteolysis potential, are a function of the bulk material as well as its debris. Following comprehensive experiences with polyethylene (PE) wear, material science is tracking two ways to minimize the risk of a particle-induced aseptic implant loosening: (i) reduction of the PE debris by a low-wearing articulation partner; and (ii) replacement of the PE by other materials. Therefore, new ceramics (e.g., ZTA, Si3N4), as well as coatings (e.g., TiN, “diamond-like” carbon) and modifications of a bulk metal (e.g., oxidizes zirconium) or cushion bearings (polyurethane, hydrogels), are currently available for total joint replacements or have been used for pre-clinical testing. This review gives a brief overview and evaluates the potential of those that have recently been published in literature.
Keywords :
Arthroplasty , WEAR , Coating , Metal surface treatment , Cushion bearing