Title of article
Retrospective lifetime estimation of failed and explanted diamond-like carbon coated hip joint balls
Author/Authors
Hauert، نويسنده , , R. and Falub، نويسنده , , C.V. and Thorwarth، نويسنده , , G. and Thorwarth، نويسنده , , K. and Affolter، نويسنده , , Ch. and Stiefel، نويسنده , , M. and Podleska، نويسنده , , L.E. and Taeger، نويسنده , , G.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2012
Pages
7
From page
3170
To page
3176
Abstract
Diamond-like carbon (DLC) coatings are known to have extremely low wear in many technical applications. The application of DLC as a coating has aimed at lowering wear and to preventing wear particle-induced osteolysis in artificial hip joints. In a medical study femoral heads coated with diamond-like amorphous carbon, a subgroup of DLC, articulating against polyethylene cups were implanted between 1993 and 1995. Within 8.5 years about half of the hip joints had to be revised due to aseptic loosening. The explanted femoral heads showed many spots of local coating delamination. Several of these explanted coated TiAlV femoral heads have been analyzed to investigate the reason for this failure. Raman analysis and X-ray photoelectron spectroscopy (XPS) depth profiling showed that the coating consists of diamond-like amorphous carbon, several Si-doped layers and an adhesion-promoting Si interlayer. Focused ion beam (FIB) transverse cuts revealed that the delamination of the coatings is caused by in vivo corrosion of the Si interlayer. Using a delamination test set-up dissolution of the silicon adhesion-promoting interlayer at a speed of more than 100 μm year−1 was measured in vitro in solutions containing proteins. Although proteins are not directly involved in the corrosion reactions, they can block existing small cracks and crevices under the coating, hindering the exchange of liquid. This results in a build-up of crevice corrosion conditions in the crack, causing a slow dissolution of the Si interlayer.
Keywords
Coating , Delamination , Diamond-like carbon , Silicon , Corrosion
Journal title
Acta Biomaterialia
Serial Year
2012
Journal title
Acta Biomaterialia
Record number
1756373
Link To Document