Title of article :
EIS diagnosis on the corrosion behavior of TiN coated NiTi surgical alloy
Author/Authors :
Yang، نويسنده , , Dazhi and Liu، نويسنده , , Chenglong and Liu، نويسنده , , Xiaopeng and Qi، نويسنده , , Min and Lin، نويسنده , , Guoqiang، نويسنده ,
Issue Information :
دوماهنامه با شماره پیاپی سال 2005
Pages :
5
From page :
417
To page :
421
Abstract :
In this work titanium nitride (TiN) coatings were deposited on NiTi surgical alloy by arc ion plating (AIP). The open circuit testing and electrochemical impedance spectroscopy (EIS) have been employed to study the corrosion performance of the TiN coated samples in Troyde’s simulated body fluid. It was found that the TiN coating showed a good corrosion resistance at the beginning of long-term immersion; with the increase of immersion time, however, the corrosion resistance of the coating deteriorated after nearly 24 h of immersion; moreover, the large areas of the coating had fallen off the substrate after 30 days of immersion.
Keywords :
NiTi surgical alloy , Titanium nitride , Electrochemical impedance spectroscopy (EIS) , CORROSION RESISTANCE , Simulated Body Fluid
Journal title :
Current Applied Physics
Serial Year :
2005
Journal title :
Current Applied Physics
Record number :
1769897
Link To Document :
بازگشت