Title of article :
The calculation and measurement of synchrotron radiation flux: finite emittance effects
Author/Authors :
Bawagan، نويسنده , , A.D.O، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Pages :
8
From page :
187
To page :
194
Abstract :
A finite emittance model (CLSxBW) for calculating the flux of synchrotron radiation is described. CLSxBW accounts explicitly for the finite source size and divergence of the electron beam in the storage ring. Finite source effects are found to be very critical for the proper calculation of the linear polarization of the synchrotron radiation. Intercomparisons of CLSxBW with Schwinger (Bessy computer code) show agreement at 1 and 10 ppm for rectangular and circular detector apertures, respectively. Comparison of emitted power calculated with CLSxBW and power measurements on the white light beamline in BessyII using a broadband detector show excellent quantitative agreement.
Journal title :
Chemical Physics Letters
Serial Year :
2001
Journal title :
Chemical Physics Letters
Record number :
1774642
Link To Document :
بازگشت