• Title of article

    Development of a recording system for saliva pH with a complete denture by telemetry

  • Author/Authors

    Watanabe، نويسنده , , Tatsutomi and Kobayashi، نويسنده , , Kenʹichi and Suzuki، نويسنده , , Tetsuya and Kumagai، نويسنده , , Hiroshi and Ichikawa، نويسنده , , Kenichi and Kobayashi، نويسنده , , Kosei and Kyu، نويسنده , , Thein and Sato، نويسنده , , Ikuko and Nagao، نويسنده , , Masanori، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1997
  • Pages
    3
  • From page
    251
  • To page
    253
  • Abstract
    Summary system for measuring saliva pH by radio telemetry was developed. The system was built into a lower complete denture and could operate for about 19 h. The maximum error in the range pH 5–8 was 0.15 pH, and the correlation when values were compared with those from a glass-electrode pH meter was 0.99. That this system operates successfully was confirmed in a clinical experiment.
  • Keywords
    PH , telemetry , Complete denture , Saliva
  • Journal title
    Archives of Oral Biology
  • Serial Year
    1997
  • Journal title
    Archives of Oral Biology
  • Record number