Title of article
Development of a recording system for saliva pH with a complete denture by telemetry
Author/Authors
Watanabe، نويسنده , , Tatsutomi and Kobayashi، نويسنده , , Kenʹichi and Suzuki، نويسنده , , Tetsuya and Kumagai، نويسنده , , Hiroshi and Ichikawa، نويسنده , , Kenichi and Kobayashi، نويسنده , , Kosei and Kyu، نويسنده , , Thein and Sato، نويسنده , , Ikuko and Nagao، نويسنده , , Masanori، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1997
Pages
3
From page
251
To page
253
Abstract
Summary
system for measuring saliva pH by radio telemetry was developed. The system was built into a lower complete denture and could operate for about 19 h. The maximum error in the range pH 5–8 was 0.15 pH, and the correlation when values were compared with those from a glass-electrode pH meter was 0.99. That this system operates successfully was confirmed in a clinical experiment.
Keywords
PH , telemetry , Complete denture , Saliva
Journal title
Archives of Oral Biology
Serial Year
1997
Journal title
Archives of Oral Biology
Record number
1800618
Link To Document