On doubling the signal in localised stimulated echo measurements
Author/Authors
Payne، نويسنده , , G.S and Leach، نويسنده , , M.O.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1995
Pages
4
From page
629
To page
632
Abstract
Stimulated echo (STEAM) localisation yields up to 50% of the full magnetisation. It is shown that give the good magnetic field homogeneity now obtained in clinical MRS, and with a slight modification to the gradient pattern, the localised signal may be nearly doubled, although at the expense of introducing a frequency-dependent amplitude for peaks in the final spectrum. A two-step phase cycle is also required. The principle is demonstrated on a clinical MR system, and the relative merits of the sequence are discussed.
Keywords
1H MRS , Stimulated echo (STEAM) , Sequence development , Localised spectroscopy
Journal title
Magnetic Resonance Imaging
Serial Year
1995
Journal title
Magnetic Resonance Imaging