Accuracy and precision of time-averaged flow as measured by nontriggered 2D phase-contrast MR angiography, a phantom evaluation
Author/Authors
Bakker، نويسنده , , C.J.G and Kouwenhoven، نويسنده , , M and Hartkamp، نويسنده , , M.J and Hoogeveen، نويسنده , , R.M and Mali، نويسنده , , WPTM Mali، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1995
Pages
7
From page
959
To page
965
Abstract
The purpose of this study was to assess the accuracy and precision of time-averaged flow as measured by nontriggered 2D PC. Mono-, bi-, and triphasic flow patterns, modelling waveforms encountered in the human vascular system, were generated by a computer-controlled flow system. Time-averaged flow velocity was measured by conventional 2D cardiac-triggered cine PC and by nontriggered 2D PC for different settings of the excitation flip angle and the velocity sensitivity. Accuracy and precision were determined by repeating the measurements (N = 6) and comparing the results against precisely known calibration values. Measurements revealed waveform-specific deviations between triggered and nontriggered acquisitions that depended on the velocity sensitivity and, more strongly, on the flip angle of the nontriggered experiment. This confirmed the theoretically predicted predominance of amplitude over phase effects. Systematic errors could be reduced by decreasing the flip angle and the velocity sensitivity, although at the expense of signal-to-noise, so that additional signal averaging was required to maintain a specified precision. The attainable accuracy appeared to be acceptable only for waveforms with a relatively low pulsatility index. The study demonstrates the feasibility of accurate and precise nontriggered velocity measurements for weakly pulsatile flow and indicates a route towards improving the reliability for highly pulsatile flow.
Keywords
Flow studies , Flow dynamics , Pulsatile flow , Magnetic resonance angiography , magnetic resonance , Phantom studies , Cardiac triggering
Journal title
Magnetic Resonance Imaging
Serial Year
1995
Journal title
Magnetic Resonance Imaging
Record number
1827099
Link To Document