• Title of article

    Evaluation of gradient inhomogeneity in the optimal design of gradient coils

  • Author/Authors

    Du، نويسنده , , Peter J. Parker، نويسنده , , Dennis L.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1996
  • Pages
    7
  • From page
    201
  • To page
    207
  • Abstract
    The image distortion caused by gradient inhomogeneity was derived from the image equation of Fourier MR imaging. A measurement of gradient inhomogeneity was proposed based on the amount of image distortion. Using the proposed measurement of gradient inhomogeneity in the optimal design of a z-gradient coil reduced image distortion by 39–45% compared to a previously used figure of merit in gradient coil design.
  • Keywords
    Gradient inhomogeneity , Optimal design , Gradient coils
  • Journal title
    Magnetic Resonance Imaging
  • Serial Year
    1996
  • Journal title
    Magnetic Resonance Imaging
  • Record number