Title of article :
An interleaved sequence for accurate and reproducible clinical measurement of Magnetization Transfer Ratio
Author/Authors :
Barker، نويسنده , , G.J. and Tofts، نويسنده , , P.S. and Gass، نويسنده , , A.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1996
Pages :
9
From page :
403
To page :
411
Abstract :
We demonstrate an interleaved dual spin echo-based sequence for quantitative measurement of Magnetisation Transfer Ratio (MTR) in a clinical environment that overcomes the problems of patient motion between scans faced by noninterleaved methods. The sequence also provides proton density and T2-weighted images, allowing direct comparison among the three contrast regimes. Phantom studies and in vivo measurements on normal controls show the sequence to be robust in normal use. The values of MTR calculated from the sequence are shown to be precise and reproducible enough to allow regional variations to be identified within and between white matter and other brain tissues.
Keywords :
Interleaving , Spin echo , magnetization transfer
Journal title :
Magnetic Resonance Imaging
Serial Year :
1996
Journal title :
Magnetic Resonance Imaging
Record number :
1827272