• Title of article

    Analysis of sinusoidal-shaped frequency-selective rf pulses

  • Author/Authors

    Hua، نويسنده , , Jianmin and Ives، نويسنده , , F.Jay، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1996
  • Pages
    7
  • From page
    533
  • To page
    539
  • Abstract
    RODEO (rotating delivery of excitation off resonance), which can be represented as an Δ ∇ pulse where Δ is an 2β-sinusoidal waveform and ∇ = −Δ, has been used as a frequency-selective rf pulse for fatsuppressed three-dimensional magnetic resonance imaging. This study systematically compared several sinusoidal-shaped pulses with different combinations of Δ and ∇. The sinusoidal-shaped pulses were also compared with the Gaussian-shaped and binomial-shaped pulses. The overall performances for fat suppression can be rated as “Δ ∇ Δ > 1331 > Δ ∇ = 1bar21”, where 121 and 1331 are second- and third-order binomial pulses, respectively.
  • Keywords
    Magnetic resonance imaging (MRI) , Fat Suppression , Frequency-selective rf pulses